Power Box (PD-TaD 80 as a standalone system)

The Power Box acts as the power supply for all system components:
PD-TaD 80
VLF HV generator
Laptop
No.
Element
1
Mains voltage
connection
2
Device protection fuse
3
Laptop port
4
PD Detector port
5
Ports for system power
supply
18 / 74
Function
Is used to connect the Power Box to the mains voltage
When the mains voltage is connected, the LED next to the mains voltage
connection lights up.
Used to protect the Power Box and connected system components
Type: T 16 AH
Is used to connect the laptop
The laptop is used for data transmission and is connected to the Power Box
via the Ethernet cable (included in the standard delivery).
Is used to connect the PD measuring unit of the PD-TaD 80
The PD measurement unit is connected to the Power Box via a PoE cable.
The PoE cable is used for the power supply of the PD-TaD 80 and for data
transmission.
Are used to connect the VLF HV generator and the laptop to the power
supply on the Power Box
Max. current 16 A
PD-TaD 80
822-188-3