Intended use
Transferring patients (adult or children) between devices (e.g., within the room), floor lifting, horizontal lifting, supporting
patient limbs, ambulating patient, bathing patient, toileting patient, weighing patient and transferring patients from car.
Intended for use in following environments: health care, intensive care, emergency ward, rehabilitation, habilitation.
This product is not intended to be used by the patient alone. Lifting and transferring a patient shall always be performed
with the assistance of at least one caregiver. This product is used as a means to perform the lift but is not in contact with the
patient; therefore we do not discuss or describe various patient conditions in this instruction for use. Contact your Hill-Rom
representative for support and advice.
Certain environments and conditions can limit the correct us of the mobile lifts, including:
Thresholds, unlevel floor surfaces, various obstacles, and extra-thick carpets. These environments and conditions can cause the
wheels of the mobile lift not to roll as intended, possible imbalance in the mobile lift, and increased exertion by the caregiver. If
you are uncertain that your care environment fulfills the requirements for correct use of the mobile lift, please contact your Hill-
Rom representative for further advice and assistance.
Unbalanced lifting poses a tipping risk and may damage the lift equipment!
Never leave a patient unattended during a lifting situation!
Before use, make sure that:
• the lift is assembled in accordance with the assembly instructions;
• the lifting accessories are properly attached to the lift;
• the battery has been charged for at least 6 hours;
• you have read the instructions for use for the lift and lifting accessories;
• personnel using the lift are informed of the correct operation and use of the lift.
Before lifting, always make sure that:
• the lifting accessory is selected appropriately, in terms of type, size, material and design, with regard to the patient's needs;
• the lifting accessories are not damaged;
• the lifting accessories are correctly attached to the lift;
• the lift strap is not twisted or worn and can move in and out of the lift;
• the lifting accessories hangs vertically and can move freely;
• the lifting accessory is correctly and safely applied to the patient in order to prevent injuries;
• the sling bar latches are intact. Missing or damaged latches must always be replaced with new ones;
• the sling's strap loops are correctly connected to the sling bar hooks when the sling straps are extended
but before the patient is lifted from the underlying surface.
Incorrect attachment of sling to slingbar may cause severe injury to the patient.
Prod. No. 2000045 and 2000049 have been tested by an accredited testing institute
No modification of this product is allowed.
Use of this equipment adjecent to other equipment should be avoided because it could result in improper operation, if such
use is neccessary, observe and verify that other equipment are operating normally.
Particular care must be observed when using strong sources of potential disturbance, such as diathermy, etc, so
that diathermy cables are not positioned on or near the lift. If you have questions, please consult the responsible
assistive device technician or the supplier.
The lift may not be used in areas where flammable mixtures may occur, for example, in areas where flammable
goods are stored.
This Caution notice is found on the Battery:
CAUTION! NOT TO BE OPENED BY UNAUTHORIZED PERSONNEL
DO NOT SHORT CIRCUIT
USE THE SPECIFIED CHARGER ONLY
MAY EXPLODE IF DISPOSED IN FIRE
This Caution notice is found on the Control box:
CAUTION! NOT TO BE OPENED BY UNAUTHORIZED PERSONNEL