Cleaning And Disinfection - Liko Golvo 9000 Series Instructions For Use Manual

Mobile lift
Table of Contents

Advertisement

Safety recommendations
Cleaning and disinfecting procedures for Liko™ Mobile lifts. This instructions do not replace the facility's own cleaning and
disinfection policies.
• Wear protective equipment according to manufacturer's instruction and per facility protocol throughout the cleaning
operations, such as: rubber gloves, goggles, apron, face mask and shoe covers.
• Unplug mains (AC power source) before cleaning and disinfection.
• Never clean the lift by pouring water on it, steam cleaning it, or by using a high-pressure jet.
• Refer to the recommendations made by the cleaning and disinfecting product manufacturer.
Equipment:
Protective equipment (such as: rubber gloves, goggles, apron, face mask and shoe covers) as recommended by the facility
protocol and manufacturers instructions
Clean buckets
Cloths for washing and drying
Soft brush
Warm water
To find Cleaning / Disinfectants compatible or not compatible for use on Liko's products, follow the
"Application of commonly used Cleaning / Disinfectants on Liko products" in this document.
Cleaning instructions
1.
Unplug mains (AC power source) before cleaning and disinfection.
2. Clean the lift with a cloth moistened with warm water and a neutral cleaning agent approved by your organization.
A soft brush may be used to remove stains and resistant dirt.
3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to all areas
run the lift into the highest and lowest positions and extend the base width adjustment entirely out. Remove the Battery to
have access behind the battery. Lower the sling bar to get access to the whole lift strap. Fully extend the lift strap by using
the emergency lowering. After cleaning the lift strap, and before you raise the sling bar, make sure the lift strap has dried.
After the emergency lowering device is used, the lifting level needs to be restored; see "Operation" in this document.
4. Pay special attention to the following areas:
• Lift strap
• Sling bar (different designs)
• Armrest
• Mechanical emergency lowering
• Hand control
• Handles
• Control box
• Emergency stop
• Battery
• Wheels
Disinfection Instructions
1. For the use of suitable disinfectants see "Application of commonly used Cleaning / Disinfectants on Liko products"
in this document.
2. Use the choice of disinfectant according to the manufacturer's instructions and repeat the work step as in
"Cleaning instructions".
3. Remove traces of disinfectant after disinfection. Wipe off the lift with a cloth moistened with clean water starting from the
top and working down. The cloth shall not be so damp that it drips.
The lift may not be cleaned with CSI or equivalent.
The hand control may not be cleaned with Viraguard or equivalent.
The control box may not be cleaned with Anioxy Spray or equivalent.
The lift strap may not be cleaned with Oxivir Tb, Dispatch, Chlor-Clean, Dismozon Pur or equivalent.
7EN140108 Rev. 4 • 2020

Cleaning and Disinfection

23
www.hillrom.com

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Golvo 9000 lowbase20000452000049

Table of Contents