Inspection And Maintenance - Liko Golvo 9000 Series Instructions For Use Manual

Mobile lift
Table of Contents

Advertisement

For trouble-free use, certain details should be checked each day the lift is used:
• Inspect the lift and check to make sure that there is no external damage.
• Check the sling bar attachment.
• Check the lift strap for wear and to ensure the strap is not twisted.
• Check the functionality of the latches.
• Check the integrity of the lifting motion and the base-width adjustment.
• Check that the lifting interval level is set correctly and that the emergency lowering is working as it should (both the elec-
trical and the mechanical).
• Charge the battery each day the lift is used and make sure the charger works.
When necessary, clean the lift with a moist cloth and check that the wheels are free from dirt. Find more detailed information
regarding cleaning and disinfection of your Liko™ product in chapter; Cleaning and Disinfection.
The lift should not be exposed to running water.
Service
A periodic inspection of the lift should be carried out at least once per year.
Periodic inspection, repair and maintenance may be performed only in accordance with the Liko service manual by
personnel authorized by Hill-Rom and using original Liko spare parts.
Service activities are not allowed with the patient in the lift.
Service Agreement
Hill-Rom offers the opportunity to enter into service contracts for the maintenance and regular inspection of your
Liko product.
Expected Life Time
The product has an expected service life of 10 years when correctly handled, serviced and periodically inspected in
accordance with Liko's instructions.
Parts listed below are subject to wear and tear and have specific expected life time:
- Handcontrol, expected life time 2 years,
- Battery, expected life time 3 years.
Transport and Storage
During transportation, or when the lift is not to be used for a long time, the emergency stop should be activated.
The environment where the lift is transported and stored should have a temperature of -10°C to +50°C (14°F to 122°F),
20-90% humid., pressure 700-1060 hPa.
The environment where batteries is transported and stored should have a temperature of -10°C to +40°C (14°F to 104°F) ,
20-80% humid., pressure 700-1060 hPa.
Product Changes
Liko products undergo continuous development, which is why we reserve the right to make product changes without prior
notice. Contact your Hill-Rom representative for advice and information about product upgrades.
Design and Quality by Liko in Sweden
The management system for both manufacturing and development of the product is certified in accordance with ISO9001
and its equivalent for the medical device industry, ISO13485. The management system is also certified in accordance with the
environmental standard ISO14001.
Notice to Users and/or Patients in EU
Any serious incident that has occurred in relation to the device, should be reported to the manufacturer and the competent
authority of the Member State in which the user and/or patient is established.
7EN140108 Rev. 4 • 2020

Inspection and Maintenance

25