Averaging

Averaging helps to reduce the effects of random noise on a measurement. The
FieldFox computes the displayed value over several consecutive readings.
Average count determines the number of readings to average. The higher the
average count, the greater the amount of noise reduction.
An average counter is shown in the left edge of the screen as Avg N. This shows
the number of previous readings that have been averaged together. When the
counter reaches the specified count, then a „running average‟ of the last N
readings is displayed.
Average count = 1 means there is NO averaging.
This setting can be changed at any time without affecting calibration accuracy.
How to set Averaging
 Press Meas Setup 4
 Then
 Then
 Enter a value using the numeric keypad. Enter 1 for NO averaging.
 Press Enter.
To restart the averaging process, press Meas Setup 4, then More, then
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
 Press Sweep 3.
 Then choose one of the following:
 You can also use Run / Hold +/- to toggle between Single and Continuous.
72
Low Sets output power to approximately –25 dBm.
o
More
Average N
where N is the current count setting.
Single
Single
Automatically sets Continuous OFF and causes the FieldFox to
o
make ONE measurement, then hold for the next Single key press. When a
data trace is displayed, the entire trace is measured, then holds. The
annotation changes to an arrow
Makes continuous measurements. This is the typical setting
Continuous
o
when battery power is not critical.
button is pressed. Use Single to conserve battery
-->
while the measurement occurs.
Restart
Hold
N9923A FieldFox User's Guide