You can also use Run / Hold +/- to toggle between Single and Continuous.
Resolution (Number of Data Points)
Data points are individual measurements that are made and plotted across the X-
axis to form a trace. Select more data points to increase measurement resolution.
However, more data points requires more time to complete an entire
measurement sweep.
When the Resolution is changed after a calibration is performed, the cal becomes
interpolated. Learn more on page 51.
How to set Resolution
 Press Sweep 3.
 Then Resolution.
 Then choose from the following:
 Using SCPI, Resolution can be set to any number of points between 3 and
Sweep Time
The fastest possible sweep time is always used as the default setting. Use the Min
Swp Time setting to slow the sweep time when measuring long lengths of cable.
Learn more about measuring long cable lengths at the FieldFox Supplemental
User‟s Guide:
The actual sweep time is shown on the FieldFox screen. See the Screen Tour on
page 17. To increase the sweep time, enter a value that is higher than the actual
sweep time. The increase will not be exactly the amount that you enter, as the
actual sweep time is the composite of many factors.
NOTE
Measurement speed specifications do NOT apply in Temperature Control Mode.
Learn more on page 13.
How to set Sweep Time
 Press Sweep 3.
 Then
 Enter a value using the numeric keypad.
 Press a multiplier key. Learn about multiplier abbreviations on page 18.
28
Automatically sets Continuous OFF and causes FieldFox to make
Single
o
ONE measurement, then hold for the next Single key press. When a data
trace is displayed, the entire trace is measured, then holds. The
annotation changes to an arrow
Continuous ON OFF
o
setting when battery power is not critical.
10001. See the Programming Guide at
http://na.tm.agilent.com/fieldfox/help/
Min Swp
Time.
-->
while the measurement occurs.
Makes continuous measurements. This is the typical
101 | 201 | 401 | 601 | 801 |
http://na.tm.agilent.com/fieldfox/help/
Hold
1001.
N9923A FieldFox User's Guide