Installation
12
5.1.2
Take viscosity into consideration
Low viscosity
8
Installation example for low-viscosity liquids. Unit of measurement mm (in)
D
Diameter of installation socket: at least 50 mm (2.0 in)
Low viscosity, e. g. water: < 2 000 mPa⋅s
It is permitted to position the tuning fork within the installation socket.
High viscosity
9
Installation example for a highly viscous liquid. Unit of measurement mm (in)
NOTICE
Highly viscous liquids may cause switching delays.
‣
Make sure that the liquid can run off the tuning fork easily.
‣
Deburr the socket surface.
High viscosity, e. g. viscous oils: < 10 000 mPa⋅s
The tuning fork must be located outside the installation socket!
> 25 (0.98)
> 40 (1.57)
Liquiphant FTL51B
A0033297
A0037348