Installation
12
5.1.2

Take viscosity into consideration

Low viscosity
 8
Installation example for low-viscosity liquids. Unit of measurement mm (in)
D
Diameter of installation socket: at least 50 mm (2.0 in)
Low viscosity, e. g. water: < 2 000 mPa⋅s
It is permitted to position the tuning fork within the installation socket.
High viscosity
 9
Installation example for a highly viscous liquid. Unit of measurement mm (in)
NOTICE
Highly viscous liquids may cause switching delays.
Make sure that the liquid can run off the tuning fork easily.
Deburr the socket surface.
High viscosity, e. g. viscous oils: < 10 000 mPa⋅s
The tuning fork must be located outside the installation socket!
> 25 (0.98)
> 40 (1.57)
Liquiphant FTL51B
A0033297
A0037348