Sending The Device In For Maintenance; Changing The Battery - Weinmann MEDUMAT Easy CPR Instructions For Use Manual

Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

10.2 Sending the device in for
maintenance
Risk of infection and contamination due to inadequate
hygienic safety during maintenance measures!
The device and accessories may be contaminated, and infect the
technicians or bystanders with bacteria or viruses.
 Carry out cleaning and disinfection of the device and
accessories prior to all maintenance.
 Never send in potentially contaminated devices or accessories
for maintenance work.
1. Remove the accessories
2. Hygienically reprocess the device and accessories
(see "8 Cleaning and disinfection", page
3. Send the device and accessories to the manufacturer or
technicians who have been expressly authorized by the
manufacturer
Result
The device has been sent in for maintenance.
If you send in devices or accessories which are visibly
contaminated, they will be disposed of by the manufacturer or a
technician specifically authorized by manufacturer at your expense.

10.3 Changing the battery

The device comes with a special 3.6 V battery.
As a rule, the battery is changed regularly within the scope of
2-yearly maintenance and, as such, if the device is used normally,
the operator does not have to change the battery.Nevertheless, it
is recommended always to keep a spare battery
(WM 28045 LSH 14) with the device
parts", page 102)
(see "7 Disassembly", page
(see "1.2 Customer Service", page
and
(see "5 Function check", page
MEDUMAT Easy
10 Maintenance
70).
79).
6).
(see "12.3 Replacement
39).
CPR
EN
97

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 20310Wm 20330

Table of Contents