4.
CAUTION! Risk of injury from improperly removed
testing bag for function check!
Remove the testing bag from the patient valve.
Check whether the device triggers the pAW
alarm after the second inspiration breath.
5. Connect the testing bag back up to the patient valve.
Result
The tidal volume and airway pressure measurement have been
checked.
5.5.5 Checking MEDUtrigger
1. Select the following settings:
Via the adjusting knob: 950 ml (Vt) at 10 bpm (frequency)
Via the pMax button with control LEDs: 45 cmH
5 Function check
CPR
MEDUMAT Easy
/
Apnea
O
2
EN