5.5 Checking device functions
Sa
5.5.1 Checking visual and audio alarm output
Requirement
In order to perform the function check comprehensively and
rapidly, check all functions in succession in the order below.
Risk of injury from improperly removed testing bag for
function check!
If the testing bag is removed improperly, the connector of the
testing bag may remain on the breathing circuit. The resulting
increase in inspiratory airway resistance can injure the patient.
When disassembling always pull the testing bag off at the
connector.
Damage to the device due to pressure blows on fittings!
Opening the oxygen cylinder valve too quickly can lead to strong
pressure blows and can damage the oxygen cylinder or the fitting.
Always open the oxygen cylinder valve slowly.
The pAW
alarm can be ignored when performing the following
testing steps unless testing the alarm itself is specifically required.
•
The device and accessories have been checked visually
"5.2 Visually checking the device and accessories", page
•
The function check is ready
check", page
41).
•
The device is switched off.
1. Open oxygen cylinder.
5 Function check
(see "5.3 Preparing for the function
CPR
MEDUMAT Easy
(see
40).
EN
43