3.3 Accessories

3.3.1 Breathing circuit
6
3-2 Breathing circuit
The ventilation gas is delivered to the patient via the breathing
circuit.The breathing circuit comprises a pressure measuring tube
(1), a ventilation hose (2) and the patient valve. Both hoses are
connected to the patient valve and the device. The breathing
circuit is designed as a disposable system which must be replaced
after each use. The pressure measuring tube conducts the pressure
on the inspiration side to the device. The pressure measuring tube,
the ventilation hose and the MEDUtrigger cable are connected
together with clips (3).
The patient valve is designed so that if the device fails,
spontaneous respiration is possible regardless of the ventilation
mode set. If necessary, fresh air can be inhaled via the spontaneous
respiration side with emergency air membrane (4).
The inspiration side (6) serves to enable a breathing system filter, a
ventilation mask or a tube to be attached to the patient valve or to
connect a testing bag to perform a function check. The expiration
side (5) serves to discharge the respiratory air to the environment
during expiration and, if necessary, to connect a PEEP valve to the
patient valve.
1
4
5
MEDUMAT Easy
3 Description
2
3
CPR
EN
19