Weinmann MEDUMAT Easy CPR Instructions For Use Manual page 42

Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

5 Function check
Result
Requirement
Result
42
EN
MEDUMAT Easy
2. Read off the oxygen cylinder pressure on the contents
gauge (1) of the pressure reducer.
3. Close the oxygen cylinder.
4. Monitor the needle on the contents gauge (1) on the pressure
reducer for 1 minute.
If the position of the needle remains constant: The system
is free from leaks.
If the needle falls continuously: The system is not leakproof.
The system has been checked for leaks.
If the system is leaking, remedy the system leak as described below.
Rectifying leaks in the system
The system is not leakproof.
1. Check that the tubes are connected correctly
checking the device and accessories", page
2. Check that the screw connections are tightened correctly
"5.2 Visually checking the device and accessories", page
3. Check the system for leaks once more
system for leaks", page
4. If a leak in the system cannot be remedied, have the device
repaired
(see "1.2 Customer Service", page
The leak in the system has been rectified.
CPR
(see "5.4 Checking the
41).
1
(see "5.2 Visually
40).
(see
40).
6).

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 20310Wm 20330

Table of Contents