5 Function check
44
EN
MEDUMAT Easy
2. Connect the testing bag for function check up to the patient
valve.
3. Switch on the device using the On/Off button.
Upon being switched on, the device performs an automatic
self-test which takes approx. 2 seconds.
4. Monitor the self-test and also check the following signals:
Check whether the device emits an audio signal when
being switched on.
4
3
2
Check that all the LEDs in the control panel light up at least
once when being switched on: Demand flow (1), pMax (2),
ventilation pressure display (3), alarm (4), alarm mute (5),
Man. (6).
CPR
pMax (cmH
O)
2
45
20
5
6
1