5 Function check
Result
52
EN
MEDUMAT Easy
Check whether the device switches the flow on and
immediately off again. You can hear a slight click.
3. Switch off the device using the On/Off button. To do this, keep
the On/Off button depressed until all 4 alarm LEDs light up.
Then release the On/Off button.
4.
CAUTION! Risk of injury from improperly removed
testing bag for function check!
Grasp the testing bag by the connector and pull the bag and
connector off the patient valve.
5. Close the oxygen cylinder.
The demand flow mode has been checked.
The function check is complete.