8 Cleaning and disinfection
84
EN
MEDUMAT Easy
2. The cleaning agents, doses, and exposure time for the
individual parts can be found in the cleaning and disinfection
plan
(see "8.2 Cleaning and disinfection plan", page
3. Prepare the cleaning solution in accordance with the
specifications of the cleaning agent manufacturer.
4.
Warning! Risk of injury and damage to the device
caused by ingress of liquids!
Do not remove the compressed gas hose from the compressed
gas connection.
5.
CAUTION! Risk of injury and damage to the device
caused by ingress of liquids!
Turn the connection for ventilation hose downward.
6. To remove all visible soiling: Thoroughly brush down parts on
the inside and outside using a standard soft lumen brush or
other brush that is suitable for use on plastics and that has
been wetted with the cleaning agent.
When doing so, note: Keep uneven surfaces and grooves (e.g.,
belt buckle and complete belt surfaces, top and bottom of
MEDUtrigger, adjusting knob, connection for ventilation hose)
moist for the duration of the exposure time and brush them off
particularly thoroughly.
7. If the parts need to be wiped down acc. to the cleaning and
disinfection plan: Wipe parts down with a clean, lint-free cloth
that has been moistened with cleaning solution.
When doing so, note: