2 Safety
2.4.2 Disposable items
2.5 Safe use of the device and accessories
2.5.1 Preventing interference between the devices
12
EN
MEDUMAT Easy
disinfection plan
page
79).
Wear suitable protective equipment (e.g., gloves) during
cleaning and disinfection.
Only use the specified products for cleaning and disinfection.
Re-using disposable items causes the following risks:
•
Infection if items come into contact with airways
•
Malfunctions when using the device
•
Unforeseeable reactions as a result of aging, embrittlement,
wear, thermal load and the effects of chemical processes
Never perform cleaning and disinfection of disposable items.
Never use disposable items more than once
Electrical devices which are operated directly next to or on top of
each other can cause mutual interference to functionality. Portable
high-frequency communication devices in the direct vicinity of the
device can also influence the functioning of the device.
Do not stack the device with other electrical devices.
Do not operate the device directly next to other electrical
devices. Exception: Other WEINMANN Emergency devices
which have been tested and shown to guarantee interference-
free operation with the adjacent device. A list of other devices
is available on request.
If stacking or operation in the immediate vicinity cannot be
avoided: Closely monitor the functioning of all affected
medical electrical devices and do not use if functions are
disrupted.
With portable RF communication devices, maintain a minimum
distance of 30 cm (approx 12 inches) to the device and
accessories. Examples: Wireless device, cell phone.
CPR
(see "8.2 Cleaning and disinfection plan",