Instructions for use SMARTmatic S80 K - 1.011.6771
5 Operation | 5.4 Removing the dental burs or diamond grinders
Hazard from rotating bur or diamond grinder.
Lacerations.
▶ Do not touch the bur or diamond grinder when it is rotating!
▶ Remove the bur/diamond grinder from the medical device after treatment to avoid
injury and infection when putting it away.
Use of dental burs or diamond grinders with worn or damaged shafts.
Risk of injury, tool may fall out during treatment.
▶ Never use dental burs or diamond grinders with damaged or worn shafts.
Hazard from defective chuck system.
The tool can fall out and cause injury.
▶ Pull on the tool to check if the chucking system is functioning properly and that
the tool is firmly clamped. Wear gloves or a thimble when you check, insert, or re-
move the bits to prevent injury and infection.
Tool shaft slips inside the chuck due to excessive speed of the tool or abrupt engage-
ment of the tool.
Material damage to tool shaft and chuck system, reduction of the service life of tool
and chuck system.
▶ Do not operate the tool at a higher speed than recommended by the manufac-
turer.
Use of dental burs or diamond grinders with worn or damaged shafts.
Material damage to the chuck system, tool is difficult or impossible to remove from the
chuck system.
▶ Do not use dental burs or diamond grinders with damaged or worn shafts.
▶ Forcefully press the push button with your thumb and simultaneously insert the
dental bur or diamond grinder all the way to the stop.
▶ Check that the dental bur or diamond grinder is securely attached by pulling on it.

5.4 Removing the dental burs or diamond grinders

Hazard from rotating bur or diamond grinder.
Lacerations and damage to the chuck system.
▶ Do not touch the bur or diamond grinder when it is rotating!
▶ Do not press the push-button while the cutter or grinder is rotating.
▶ Remove the cutter/grinder from the contra-angle handpiece after treatment to
avoid injury and infection when putting it away.
▶ After the dental bur or diamond grinder has stopped rotating, firmly press the
press-button with your thumb and simultaneously pull out the dental bur or dia-
mond grinder.
WARNING
CAUTION
CAUTION
NOTICE
NOTICE
WARNING
12 / 24