• Press Next to go to the next menu:
1.8
Advised sensor position
34.30 mm
Signal quality
< Previous | Next >
< Previous | Next >
< Previous | Next >
< Previous | Next >
13-04-2010 14:11:09
Advised distance [mm]
< 190
≥ 190
INFORMATION!
The maximum distance that can be covered with 1 rail is 195 mm.
The minimum distance for two rails is 180 mm.
INFORMATION!
On request it is possible to use the large rail of the OPTISONIC 6300 with
0,5 MHz transducers. Using this you can measure up to DN4000.

3.4 Step 3: Mount the sensor rails

Before mounting the rails, determine the colors on the connectors of the transducers. Make
sure that the blue transducer is upstream and the green transducer is downstream.
Installation with metal straps (DN15...250)
Put the metal straps around the pipe. Put the sensor rail(s) on the pipe including the transducers
with fixed cables.
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
INSTALLATION FOR FLOW MEASUREMENT
0%
13 MB free
www.krohne.com
Number of rails needed
1
2
3
23