INSTALLATION
3
3.4 Step 3: Mount the sensor rails
Before mounting the rails, determine the colors on the connectors of the transducers. Make
sure that the blue transducer is upstream and the green transducer is downstream. Consult the
manual for more information
Installation with metal straps (DN15...250)
Put the metal straps around the pipe. Put the sensor rail(s) on the pipe including the transducers
with fixed cables.