Time
Let the date and time on smart device automatically update when
traveling across time zones or switch between standard and daylight-
saving times.
Don't manually change smart device time. It can make the time on the
trend screen wrong and the app may stop displaying data.
Follow G6 Pro instructions. If not, patient could have a severe low or
high glucose event.
Inspect
Don't use a damaged or cracked transmitter. A damaged transmitter
could cause injuries from electrical shocks and may make the G6 Pro
not work correctly.
Use as Directed
The transmitter is small part and poses a choking hazard, particularly
for children.
Use Meter During Sensor Warmup
When a new sensor is started, there won't be any G6 Pro readings or
alarm/alerts. Use meter to make treatment decisions during the 2-hour
sensor warmup period.
Follow G6 Pro instructions. If not, patient could have a severe low or
high glucose event.
Wire Breaks Off
Don't ignore broken or detached sensor wires. A sensor wire could
remain under the skin. If this happens, please contact our 24/7
Technical Support.
If a sensor wire breaks off under the skin and your patient can't see
it, tell them not to remove it, but to contact you or seek professional
medical help. Advise your patient to seek professional medical help if
they have symptoms of infection or inflammation – redness, swelling,
or pain – at the insertion site.