Trend Accuracy
Trend accuracy explains how well the G6 captures the time-
dependent characteristics of glucose fluctuation.
The following examples quantify G6's Trend Accuracy:
1. When the G6 CGM rate of change is rapidly rising (≥ 2 mg/dL/min),
how often is reference glucose also rising? The answer is 71.3% of
the time for adults and 67.1% for pediatrics.
2. When the G6 CGM rate of change is rapidly falling (≤ 2 mg/dL/ min),
how often is reference glucose also falling? The answer is 98.0% of
the time.
3. When the G6 CGM rate of change is stable (≥ -1 mg/dL/min and ≤ 1
mg/dl/ min), how often is glucose changing rapidly (≥ 2 mg/dL/min
or ≤ 2 mg/dL/ min)? The answer is only 1.9% of the time.
Table 4-A� Trend Accuracy Rate of Change (Adults; n=159)
CGM Rate
Range
(mg/dL/
min)
<-2
[-2,-1)
[-1,0)
[0,1]
(1,2]
>2
G6 Pro User Guide
Appendix G: Technical Information
YSI glucose range (mg/dL)
<-2
[-2,-1)
53.3%
35.0%
7.4%
56.9%
0.4%
9.5%
0.1%
1.0%
0.0%
0.4%
0.1%
0.1%
[-1,-0)
[0,1]
9.9%
1.5%
32.5%
2.9%
76.9%
12.5%
26.2%
60.6%
3.1%
26.8%
0.8%
5.6%
(1,2]
>2
0.0%
0.2%
0.3%
0.0%
0.6%
0.1%
10.6%
1.6%
52.9%
16.8%
22.1%
71.3%
CGM-YSI
Pairs
(n)
463
2,077
7,986
5,199
1,734
1,367
127