Table 1-B� G6 Accuracy to YSI within CGM Glucose Ranges
(Adults; n=159)
CGM
Number
Glucose
of paired
Range
CGM-
1
(mg/dL)
YSI
<54
383
54-69
1,537
70-180
9,453
181-250
4,093
>250
3,863
CGM readings are within 40-400 mg/dL, inclusive.
1
Table 1-C� G6 Accuracy to YSI within CGM Glucose Ranges
(Pediatrics*; n=165)
CGM
Number
Glucose
of paired
Range
CGM-
1
(mg/dL)
YSI
<54
90
54-69
262
70-180
3,144
181-250
1,360
>250
916
* Includes pediatric subjects 6-17 years of age; no YSI measurements
were taken for pediatric subjects 2-5 years of age.
CGM readings are within 40-400 mg/dL, inclusive.
1
G6 Pro User Guide
Appendix G: Technical Information
Percent
Percent
within
within
15 mg/
20 mg/
dL YSI
dL YSI
84.3
90.6
89.6
95.1
Percent
Percent
within
within
15 mg/
20 mg/
dL YSI
dL YSI
48.9
62.2
85.5
88.5
Percent
Percent
within
within
40 mg/
15%
dL YSI
YSI
98.4
99.5
73.9
80.2
91.1
Percent
Percent
within
within
40 mg/
15%
dL YSI
YSI
85.6
96.6
79.8
83.4
89.3
Percent
Percent
within
within
20%
40%
YSI
YSI
86.6
99.3
92.1
99.9
97.7
100.0
Percent
Percent
within
within
20%
40%
YSI
YSI
90.8
99.5
93.5
99.9
95.9
99.9
Mean
bias
MARD
(mg/
(%)
dL)
-6.9
13.8
-0.5
11.5
-2.8
10.9
-10.0
9.3
-3.8
7.1
Mean
bias
MARD
(mg/
(%)
dL)
-20.0
26.0
-5.9
13.3
-0.3
9.7
-1.2
8.9
9.2
7.4