Dexcom G6 Pro User Manual page 118

Continuous glucose monitoring system
Hide thumbs Also See for G6 Pro:
Table of Contents

Advertisement

Table 1-A� G6 Accuracy to YSI (n=324)
Patient
Population
Overall
Adults
(18+ YO)
Pediatrics
(6-17 YO)
Pediatrics
(2-5 YO)*
* No YSI measurements were taken for this age group; results
presented are from in-clinic CGM-SMBG matched paired
measurements.
CGM readings are within 40-400 mg/dL, inclusive.
1
G6 Pro User Guide
Appendix G: Technical Information
Total
Number
number
of
of paired
subjects
CGM-YSI
324
25,101
159
19,329
165
5,772
8
Percent
within
20/20%
YSI
% (95% LB)
91.7 (90.6)
91.6 (90.3)
92.0 (89.8)
82
92.7 (86.6)
Day 1
Percent
MARD
within
(%)
20/20%
YSI
87.8
9.8
87.1
9.9
90.2
9.6
NA
9.9
116

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the G6 Pro and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents