Number of Readings Provided

The G6 is capable of providing a reading every 5 minutes, or up to
288 readings per day. For a variety of reasons, the G6 may not display
a glucose reading and readings are "skipped." The percentage of
readings you can expect to receive from the G6 over the sensor life is
98.6%. More than 97% of the sensors captured readings at least 90%
of the time. For the G6 with auto-applicator, approximately 99% of the
sensors displayed reading every 5 minutes at least 90% of the time.
Table 8 below describes the reading captured rate by each wear day
over the sensor life.
Table 8� Reading Capture Rate by Wear Day (n=374)
Wear Day
1
2
3
4
5
6
7
8
9
10
G6 Pro User Guide
Appendix G: Technical Information
Number of Sensors
374
368
364
354
348
338
329
314
303
290
Capture Rate (%)
97.6
98.6
98.7
98.6
98.5
98.5
98.2
97.8
97.0
96.4