Dexcom G6 Pro User Manual page 120

Continuous glucose monitoring system
Hide thumbs Also See for G6 Pro:
Table of Contents

Advertisement

Table 1-D� G6 Accuracy to YSI within CGM Glucose Ranges
(Pediatrics*, Abdomen; n=99)
CGM
Number
Glucose
of paired
Range
CGM-
1
(mg/dL)
YSI
<54
60
54-69
177
70-180
1,910
181-250
775
>250
574
* Includes pediatric subjects 6-17 years of age; no YSI measurements
were taken for pediatric subjects 2-5 years of age.
CGM readings are within 40-400 mg/dL, inclusive.
1
Table 1-E� G6 Accuracy to YSI within CGM Glucose Ranges
(Pediatrics*, Buttocks; n=66)
CGM
Number
Glucose
of paired
Range
CGM-
1
(mg/dL)
YSI
<54
30
54-69
85
70-180
1,234
181-250
585
>250
342
* Includes pediatric subjects 6-17 years of age; no YSI measurements
were taken for pediatric subjects 2-5 years of age.
CGM readings are within 40-400 mg/dL, inclusive.
1
G6 Pro User Guide
Appendix G: Technical Information
Percent
Percent
within
within
15 mg/
20 mg/
dL YSI
dL YSI
40.0
51.7
87.0
88.1
Percent
Percent
within
within
15 mg/
20 mg/
dL YSI
dL YSI
66.7
83.3
82.4
89.4
Percent
Percent
within
within
40 mg/
15%
dL YSI
YSI
80.0
96.0
80.5
81.9
89.2
Percent
Percent
within
within
40 mg/
15%
dL YSI
YSI
96.7
97.6
78.8
85.3
89.5
Percent
Percent
within
within
20%
40%
YSI
YSI
91.0
99.5
95.0
100.0
96.5
99.8
Percent
Percent
within
within
20%
40%
YSI
YSI
90.4
99.4
91.6
99.8
94.7
100.0
Mean
bias
MARD
(mg/
(%)
dL)
-24.1
28.9
-6.3
13.4
-1.1
9.7
-2.3
9.1
8.0
7.5
Mean
bias
MARD
(mg/
(%)
dL)
-11.7
20.1
-5.2
13.2
0.9
9.7
0.1
8.5
11.1
7.3
118