Table 4-B� Trend Accuracy Rate of Change (Pediatrics*; n=165)
CGM Rate
Range
(mg/dL/
min)
<-2
[-2,-1)
[-1,0)
[0,1]
(1,2]
>2
* Includes pediatric subjects 6-17 years of age; no YSI measurements
were taken for pediatric subjects 2-5 years of age.

Hypoglycemia and Hyperglycemia Alerts

Low and High Glucose Alerts

The ability of the G6 to detect high and low glucose levels is assessed
by comparing G6 results to YSI results at low and high blood glucose
levels and determining if the alert may have sounded. The G6 and YSI
values were compared by pairing the G6 Pro reading and the YSI value
within before or after 15 minutes of each other. We suggest that you
ask your doctor what alert settings would be best for you.
G6 Pro User Guide
Appendix G: Technical Information
YSI glucose range (mg/dL)
<-2
[-2,-1)
47.9%
37.0%
6.6%
55.5%
0.5%
8.9%
0.0%
0.8%
0.0%
0.4%
0.0%
0.5%
[-1,-0)
[0,1]
12.8%
1.9%
33.8%
3.4%
73.7%
15.8%
25.5%
62.9%
4.4%
35.9%
1.7%
7.1%
(1,2]
>2
0.0%
0.5%
0.6%
0.1%
1.0%
0.0%
10.0%
0.8%
48.0%
11.4%
23.6%
67.1%
CGM-YSI
Pairs
(n)
211
686
2,048
1,666
546
423