Section 6
Cleaning and Decontamination
Bottles & Caps
Oven Interior/Shaking
Platform

Rotisserie

6-2
Shake 'n Stack
The hybridisation bottles and caps can be decontaminated by the following
procedure:
1. Soak the bottles and caps in a dilute detergent solution (Decon 90)
overnight.
2. Remove the Decon 90 and rinse with distilled water.
3. If the bottles and caps have been decontaminated, no further action is
necessary. If, however, they are still contaminated, gently scrub them
with an abrasive cloth or brush and if necessary, leave to decontaminate
in Decon 90 for a longer time period.
Stainless steel drip trays are provided with all Thermo ovens and are
intended to contain spillages in the event of an accident. These, together
with the stainless steel surfaces of the ovens and the shaking platform, can
be decontaminated by wiping clean with Decon 90, followed by distilled
water. No further action should be necessary.
The rotisserie may be cleaned with Decon 90 and distilled water. No
further action should be necessary. Avoid alcohol or other organic
solutions, which may affect the plastic surfaces.
Caution Before using any cleaning or decontamination method except
those recommended by the manufacturer, users should check with the
manufacturer that the proposed method will not damage the equipment.