Starting the battery test
The battery test can only be started if the device is connected to the mains power
supply.
1. Touch the Start button (6).
2. Confirm with the rotary knob.
✓ The respective battery test is started. The duration of the battery test is
decremented in hours and displayed in field (9). The result of the battery test is
displayed after completion.
If a battery test fails, the device will cancel the test.
Canceling the battery test
● Touch the Cancel button (7) and confirm.
✓ The appropriate battery test will be canceled.
8
Technical data
8.1
Ambient conditions
During operation
Temperature
Ambient pressure
Altitude
Relative humidity
During storage and transportation
Ambient pressure
Relative humidity
Temperature
Device without power supply unit or with
power supply unit PS500 (LFP)
For charging subsequent to storage
For storage up to 6 months
For storage up to 1 month
For storage up to 1 week
Device with power supply unit PS500
(VRLA)
For charging subsequent to storage
For storage up to 6 months
For storage up to 3 months
|
Supplement
Evita V series / Babylog VN series
10 to 40 °C (50 to 104 °F)
700 to 1060 hPa
Up to 3000 m (9842 ft)
10 to 90 %, without condensation formation
500 to 1060 hPa
5 to 95 %, without condensate formation
–20 to <45 °C (–4 to <113 °F)
–20 to <55 °C (–4 to <131 °F)
–20 to 60 °C (–4 to 140 °F)
–15 to 25 °C (5 to 77 °F)
–15 to 40 °C (5 to 104 °F)
Technical data
11