Dräger Evita V Series Supplement To The Instructions For Use page 13

Battery
Hide thumbs Also See for Evita V Series:
Table of Contents

Advertisement

8.2
Operating characteristics (continued)
Typical ventilation
Ventilation mode
FiO
2
PEEP
Pinsp
RR
MV (measured value)
Ambient temperature
Internal battery of ventilation unit (without
PS500)
Type
F3 fuse of the ventilation unit
Capacity
Voltage
Electric current
Battery runtime if mains power supply is
not available
Without GS500
With GS500
Charging
Charging time (to charge battery
fully)
Batteries of the power supply unit PS500
(VRLA)
Type
F1 and F2 fuse at PS500 behind the
side part
Capacity
Voltage
Electric current
Battery runtime if mains power supply is
not available
Without GS500
|
Supplement
Evita V series / Babylog VN series
The charging time applies to new and com-
pletely discharged batteries during typical
ventilation and the GS500 is not used. The
actual charging time depends on the bat-
tery charge.
If the GS500 is operating or the ambient
temperature is high, the battery charging
process may be restricted or interrupted.
PC-AC
21 Vol%
5 mbar (or hPa or cmH
20 mbar (or hPa or cmH
12 /min
6 L/min
22 °C (71.6 °F)
NiMH battery, sealed
F15A 80 V UL248
2.5 Ah
24 V
0 to 15 A
30 minutes
15 minutes
<4 hours
VRLA batteries
F15A 80 V UL248
24 Ah
24 V
0 to 15 A
240 minutes
Technical data
O)
2
O)
2
13

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babylog vn series