D4K3560 Interconnect Drawing (P/N: 020-101087-XX)
Solution
Check the MINISAS cables between the HIP and the formatters are connected to the correct colors.
Colors are correct but the Formatter Mismatch
alarm occurs
One or more Formatter Mismatch errors appear in the Status tab of the TPC, for example:
Formatter - Green HW Version
Formatter - Green HW Version
Details
One or more temperature harnesses are incorrectly connected to a formatter board or the HIP. Or,
the wrong temperature harness is being used for a particular formatter board.
For more information, refer to the following:
•
D4K Status System Technical Reference (P/N: 020-101259-XX)
•
Mirage 4K Status System Technical Reference (P/N: 020-101450-XX)
•
D4K2560 Interconnect Drawing (P/N: 020-101088-XX)
•
D4K3560 Interconnect Drawing (P/N: 020-101087-XX)
Solution
Replace the temperature harness on the formatter reporting the error.
Formatter
Temperature Harness
Red
001-112061-XX
Green
001-112062-XX
Blue
001-112063-XX
DisplayPort sources require reset when the
projector is power cycled
When the projector is power cycled, the DisplayPort source must be restarted.
D4K and Mirage 4K Troubleshooting Technical Reference
020-101452-03 Rev. 1 (10-2014)
D4K and Mirage 4K Troubleshooting
Mismatch Blue
Mismatch - Unknown
7