Fastening, cable entry

3.3.1 Fastening

Please observe illustration # 9800373-01_00 on the subsequent page.
• Fastening holes can be drilled into the plastic side pieces of the enclosure (see hatch-
ing). From the enclosure bottom in the area between 8 mm and the calculated value fas-
tening T or from the finished wall surface at a depth greater than 62 mm (see cross sec-
tion illustration).
• Keep the area clear where the adapter plates of the hinges are positioned.
• If fastening holes are required behind the mounting plate, the holes must be counter
sunk to accommodate a counter sunk screw.
• The enclosure bottom plate is equipped with 4 rivet nuts M4x6x11.5 mm. The mounting
plate rests on these rivet nuts and needs to be fastened using a spacer between the
mounting plate and the mounting screw M4x10 mm. Do NOT place the spacer between
the enclosure bottom plate and the mounting plate!

3.3.2 Cable entry

Please observe illustration # 9800373-01_00 on the subsequent page.
• Provisions for cable entry are made as per customer specification. Generally, they are lo-
cated at the top of the side panels. (see cross section illustration below).
• If there are no requests from the customer, additional cable entry holes can be drilled
into the enclosure. From the enclosure bottom in the area between 8 mm and the calcu-
lated value fastening D or from the finished wall surface at a depth greater than 62 mm
(see cross section illustration).
COMTRAXX® CP9xx-x – Control Panel