Replacing The Filter; Replacing Uv Lamp; Photo Catalytic Filter - Surgically Clean Air SCA7000E Operational Manual

Table of Contents

Advertisement

Chapter 6
CARE AND MAINTENANCE
Note:
Remove plastic packaging
from the VOC filter before
operation.
WARNING
• DO NOT reuse or clean
the VOC filter with liquid,
especially alcohol,
gasoline, paint thinner,
etc. It may cause fire!
• DO NOT try to place the
VOC filter in front of the
Electronic cell! It may
cause fire!
Reset the filter life
1. After replacing the VOC
filter, hold the TIMER button
2 seconds to reset the time
for the VOC filter
2. After replacing the UV
lamps, hold the UV LAMP
button 2 seconds to reset
the time for the UV lamps.

Replacing the Filter

When the "
"indicator light on the control panel flash, the VOC filter
needs to be replaced.
Replacing the VOC Filter
Stop the operation and unplug the unit.
6-5 Take the Electronic cell out first.

Replacing UV Lamp

When the "
" indicator light on the control panel flashes, it is time to
replace the UV lamps.
Removing the UV lamp
6-7 Remove the UV lamp blocker.
WARNING:
• Never look directly at UV lamps.
Direct exposure to germicidal light can cause temporary or chronic
damage to your eyesight, or even blindness.
• Always unplug the unit before cleaning, servicing or replacing the UV
lamps.
• Do not dispose of the UV bulb with normal household waste, hand it in
at an official collection point for recycling.

Photo Catalytic Filter

Photo catalytic filter is permanent, never needs replacing. When there is
dust on it, clean the dust off using a vacuum cleaner dusting brush. Do
not wash it in water
6-5 Hold the VOC filter's handle,
pull it toward to you.
6-8
Hold the UV lamp, pull it up
remove it from UV lamp holder.