TECHNICAL DESCRIPTION

WSK-1011 complies with the Electromagnetic Disturbances standard, IEC60601-1-2:2014.
As a medical electrical equipment, special precautions regarding the electromagnetic disturbances shall be taken at usage of the
device according to the information provided below.
• The device is not intended for use in environments where the intensity of electromagnetic disturbance is high, such as near active HF
surgical equipment and MRI (magnetic resonance imaging) equipment etc.
• Use of the device adjacent to or stacked with other equipment must be avoided because it could result in improper operation.
• Use of accessories other than those specifi ed or provided by the manufacturer could result in increased electromagnetic emissions
or decreased electromagnetic immunity of the device and result in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used at
least 30cm away from any part of the device, including specifi ed cables. Otherwise, degradation of the performance of this equipment
could result.
Please contact your dealer or the manufacturer for specifi c information regarding the compliance to the standard.