TECHNICAL DESCRIPTION

DSK-1031 complies with the EMC, electromagnetic compatibility, standard, IEC60601-1-2. Refer to the following tables for specifi c information regarding compliance to
the standard.
DSK-1031, as a medical electrical equipment, needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information
provided below.
• Portable and mobile RF communications equipments can aff ect the device.
• The use of accessories other than those specifi ed in this manual may result in increased emissions or decreased immunity of the device.
• DSK-1031 should not be used adjacent to or stacked with other equipment.
DSK-1031 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of DSK-1031 should assure that it is used in such
an environment.
Emissions test
RF emissions CISPR11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2 Class A
Voltage fl uctuations/fl icker
emissions IEC 61000-3-3
Table 201. Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Electromagnetic environment – guidance
Group 1
DSK-1031 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
Class B
DSK-1031 is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Complies