TECHNICAL DESCRIPTION

WSK-1011 complies with the EMC, electromagnetic compatibility, standard, IEC60601-1-2. Refer to the following tables for specifi c
information regarding compliance to the standard.
WSK-1011, as a medical electrical equipment, needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.
• Portable and mobile RF communications equipments can aff ect the device.
• The use of accessories other than those specifi ed in this manual may result in increased emissions or decreased immunity of the device.
• WSK-1011 should not be used adjacent to or stacked with other equipment.
Table 201. Guidance and manufacturer's declaration – electromagnetic emissions
WSK-1011 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of WSK-1011 should assure that it
is used in such an environment.
Emissions test
RF emissions CISPR11
RF emissions CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fl uctuations/fl icker
emissions IEC 61000-3-3
Compliance
Electromagnetic environment – guidance
Group 1
WSK-1011 uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
Class B
WSK-1011 is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies
Class A
buildings used for domestic purposes.
Complies
88