Nissei DSK-1031 Instruction Manual
Hide thumbs Also See for DSK-1031:

Advertisement

DSK-1031
INSTRUCTOIN MANUAL
DIGITAL BLOOD PRESSURE MONITOR DSK-1031
INSTRUKCJA OBSŁUGI
CIŚNIENIOMIERZA CYFROWEGO DSK-1031
РЪКОВОДСТВО ЗА ЕКСПЛОАТАЦИЯ
НА ЦИФРОВ ТОНОМЕТЪР DSK-1031
UŽIVATELSKÝ NÁVOD
K DIGITÁLNÍMU TONOMETRU DSK-1031
A DSK-1031 TÍPUSÚ DIGITÁLIS TONOMÉTER
HASZNÁLATI UTASÍTÁSA
MANUAL DE UTILIZARE
A TENSIOMETRULUI DIGITAL DSK-1031

Advertisement

Table of Contents
loading

Summary of Contents for Nissei DSK-1031

  • Page 1 DSK-1031 INSTRUCTOIN MANUAL DIGITAL BLOOD PRESSURE MONITOR DSK-1031 INSTRUKCJA OBSŁUGI CIŚNIENIOMIERZA CYFROWEGO DSK-1031 РЪКОВОДСТВО ЗА ЕКСПЛОАТАЦИЯ НА ЦИФРОВ ТОНОМЕТЪР DSK-1031 UŽIVATELSKÝ NÁVOD K DIGITÁLNÍMU TONOMETRU DSK-1031 A DSK-1031 TÍPUSÚ DIGITÁLIS TONOMÉTER HASZNÁLATI UTASÍTÁSA MANUAL DE UTILIZARE A TENSIOMETRULUI DIGITAL DSK-1031...
  • Page 2: Parts And Components

    This Instruction Manual is designed to assist the user with safe and eff ective operation of the digital blood pressure monitor DSK-1031 (hereinafter – the “Device”). The device is designed to measure systolic and diastolic blood pressure and heart rate reading in patients aged 12 years and older.
  • Page 3: Complete Set

    Touch control – is control of the device by slight touching it. an increase in pressure values. Therefore, blood pressure measured at home is often diff erent from that measured in the clinic. Since blood pressure increases at low tem- peratures, measurements should be made at room temperature (about 20°C).
  • Page 4: Setting Date And Time

    urement should be carried out in a quiet comfortable atmosphere at room tem- Battery Replacement Indicator perature. Exclude meal an hour before the measurement, and exclude smoking, soft Replace all the batteries when the battery replacement indicator is fl ashing on drinks, and alcohol 1.5-2 hours before the measurement.
  • Page 5: Correct Position During Measurement

    Date and time are set in the following order: year, month, day, hour and minute. CUFF PREPARATION 1. Setting the Year Use the button to increase and button to decrease the year value. Press the SET 1. Apply the cuff to your left upper arm so that the air button to confi rm and to pass to the next step.
  • Page 6: Measurement Procedure

    MEASUREMENT PROCEDURE Automatic Additional Infl ation. If on starting the measurement, the operating pressure appears to be insuffi cient due to IMPORTANT! The device has touch-sensitive buttons that hand movements or due to excessive tension in a hand, the cuff will be fi lled with air again are pressed by slight touching.
  • Page 7 Normal pulse INDICATION OF ARRHYTHMIA Systolic blood pressure continues to increase with aging, while diastolic blood pressure tends The fl ashing symbol « », appeared on the to decrease since the age of about 50 years. display reports on an irregular heartbeat It is considered that a high pulse pressure is rhythm (Fig.
  • Page 8: Information About Errors

    3. Memory unit number, measurement date and time will be displayed one by one Maximum allowable pressure: Do not talk and do not move during in the upper right corner of the display. pressure can not be measured due the measurement. 4.
  • Page 9: Warranty

    Electric supply type Four 1.5 volt LR6 (AA alkaline) batteries or AC adaptor You touched the START/STOP button Turn off the device by pressing the when installing the batteries. START/STOP button and perform the Power Consumption , W measurement again. Memory 2 units, 60 values per each + average value ADP-W5 Power Source...
  • Page 10: Care, Storage, Repair And Disposal

    CARE, STORAGE, REPAIR AND DISPOSAL CERTIFICATION AND STATE REGISTRATION 1. This device should be protected from excessive moisture, direct sunlight, strokes, The production of devices is certifi ed pursuant to international standards such as and vibration. THE DEVICE IS NOT WATERPROOF. ISO 9001, ISO 13485, ISO 14001:2004.
  • Page 11: Technical Description

    • DSK-1031 should not be used adjacent to or stacked with other equipment. Table 201. Guidance and manufacturer’s declaration - electromagnetic emissions DSK-1031 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of DSK-1031 should assure that it is used in such an environment.
  • Page 12 Table 202. Guidance and manufacturer’s declaration - electromagnetic immunity DSK-1031 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of DSK-1031 should assure that it is used in such an environ- ment. Immunity test...
  • Page 13 Table 204. Guidance and manufacturer’s declaration - electromagnetic immunity DSK-1031 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of DSK-1031 should assure that it is used in such an envi- ronment. Immunity test...
  • Page 14 Table 206. Recommended separation distances between portable and mobile RF communications equipment and DSK-1031 DSK-1031 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of DSK-1031 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and DSK-1031 as recom- mended below, according to the maximum output power of the communications equipment.
  • Page 15 NIHON SEIMITSU SOKKI CO., LTD. Nissei Healthcare (UK) Ltd. Rede House, New Barn Lane Henfield West Sussex BN5 9SJ UK 2508-13 Nakago Shibukawa Gunma 377-0293 Japan Copyright Nihon Seimitsu Sokki Co., Ltd., 2014. E473/1602/04 Registered Trade Mark.