Operating; Operation Via Key And Display - Siemens SITRANS F Series Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Operating

7.1

Operation via key and display

The meter is designed with a single key and a symbolic display for optimal dialog.
Display
Display is divided into 3 areas.
Figure 7-1
● Top area with symbols for status information.
● Middle area with actual information.
● Bottom area with index for actual information and selected menu.
Some of the information has additional information connected and display will automatically
toggle between information, see menu overview (Page 64). If key is not pressed for 10
minutes, display will time-out and return to default configured operator menu.
Key
There are three different ways the interface key will respond to being pressed:
1. A brief press (less than 2 seconds) will advance screen to next index or menu.
2. A short press (2 to 5 seconds) will enter a menu or escape menu selection.
3. A long press (more than 5 seconds) while in the operator menu (
selected value (e.g., call-up function) indicated by an "r".
A flashing "r" indicates a reset. A request for time and date setup is shown during power-up.
SITRANS FM MAG 8000
Operating Instructions, 03/2020, A5E03071515-AH
Display
7
) will activate a reset of
59