8.4

Battery replacement

Replacing battery
1. Loosen screws on transmitter top.
2. Remove transmitter top using a screwdriver.
3. Dispose of silica gel bag.
4. Replace O-ring to ensure continued IP68 enclosure rating.
– Check O-ring for damage or deformity.
– Smear O-ring with acid-free lubricating gel.
5. Push locking tab and loosen strip.
6. Remove battery pack with power still connected.
SITRANS FM MAG 8000
Operating Instructions, 03/2020, A5E03071515-AH
Service and maintenance
8.4 Battery replacement