Lamp replacement and calibration
Hide thumbs Also See for Corona process:
Table of Contents

Advertisement

Sensor calibration
36
6
Corona process alignment
Caution
SENSITIVE COMPONENTS
Do not touch the white surface of the white standard with your fingers.
Protect the white standard from dust and impurities. Cover it up with
the delivered lid when not using it.
We recommend certifying the white standard once a year.
Place the white standard in the standard holder at position P6.

Fig. 35 Mounting the white standard

Align the Corona process with the adjustment screws
"Fig. 37 Positioning the white standard" on page 37). Ensure that the light
beam is shining on the center of the white standard. Click Continue after
aligning the Corona process.

Fig. 36 Confirming the alignment

This procedure may take a few seconds.
Lamp replacement and calibration - Corona process | Service manual |
Lamp replacement and calibration Corona process
(see
6
EN_40_070_0010I