Reflex Test Screen - Interacoustics AT235 Instructions For Use Manual

Hide thumbs Also See for AT235:
Table of Contents

Advertisement

2.16.5.2 Reflex test screen

The upper bar of the sof t keys indicates the f unction in auto mode, while the lower bar shows the f unction f or
the sof t keys in manual mode.
D-0128073-C – 2022/05
AT23 5 - Instructions for Use - US
O
Activating manual mode in the ref lex test allows single ref lex measurements at
a time and optionally the pressure at which the ref lex is measured can be
manually set (see T).
P
The level is showing the ref lex activator intensity of the currently selected ref lex
measurement (Q).
Q
The pressure slider gives an indication f or at which pressure ref lex
measurements are set to be tested (in manual mode (O) only). The slider is
moved by holding the pressure button (see T) and turning the wheel.
R
The compliance meter shows an indication f or the current non compensated
compliance value and might be uses as a help to set the pressure at peak
pressure or at an of fset f rom peak pressure (in manual mode (O) only).
S
The currently selected ref lex measurement is indicated by the rectangle around
it. Within the ref lex graph also the numerical def lection value is displayed.
T
Pressing pressure allows setting the pressure manually (see Q) (in manual
mode (O) only).
Press Exclude to exclude the highlighted test. Once excluded press include to
have it back as part of the measure.
U
Pressing the arrow up button moves the ref lex selection to the previous ref lex
row. Moving the selection sideways is done with the wheel (19).
Page 24