Operating Instructions - 3D Systems VIDAR AdvantagePRO Series User Manual

For use with windows xp & windows 7
Table of Contents

Advertisement

Operating Instructions

19196-001 Rev. D
1. Turn on the film digitizer and wait for all three LEDs to turn green.
Note: Always turn on the film digitizer before turning on the
computer. This enables the computer to recognize the digitizer.
Note: Several minutes are required for the digitizer to carry out
its internal diagnostics and establish communications with the
scanning application. These activities must be completed
before you can load film into the digitizer.
2. Turn on the computer.
3. Launch the scanning application.
IMPORTANT: Remove stickers, tape, staples, paper clips, etc.
from films before scanning. These may cause serious film
feeding problems. Failure to remove extraneous items from
films will void your warranty.
CAUTION: X-ray images displayed on a computer monitor are
representative only. A variety of factors influence image density
and size—including monitor luminance and calibration, window
and level settings, zoom level, and dimensional and grayscale
inaccuracies resulting from build-up of tolerances in the
digitizer, display board and computer. For this reason, special
precautions must be exercised (e.g., calibrating the software
ruler and computer monitor) when taking measurements from
digitized images. Please refer to the user's manual for your
digitizing software for more information.
4. Hold the stack in front of you as
you would view it on a light box.
5. Slide the stack into the feeder,
between the guides. Do not overfill
the feeder:
The 50 sheet feeder can hold up
to 50 sheets at a time.
VIDAR Advantage Series User's Guide