Cleaning the Diffuser
19196-001 Rev. D
All Advantage Models
If you notice streaks (vertical artifacts) in images, you should clean the diffuser.
1. Turn off the digitizer. Allow 10 to 15 minutes for the bulb cartridge to cool to a
comfortable temperature before removing it.
2. Open the door on the side of the digitizer base:
3. Remove the lamp Cartridge Clamp using a #2 Philips screwdriver. (May not be
applicable in all models)
VIDAR Advantage Series User's Guide
Note: The door has a "door
open" sensor. When the
door is open, all processes—
including calibration and
scanning—will stop.
24