About Films - 3D Systems VIDAR AdvantagePRO Series User Manual

For use with windows xp & windows 7
Table of Contents

Advertisement

About Films

18cm long x 24cm wide film stack
or
8" long x 10" wide film stack
19196-001 Rev. D
The film digitizer handles standard 18cm x 24cm and 24cm x 30cm mammography films.
You can feed multiple films (up to 50), and you can mix film sizes.
IMPORTANT: Remove stickers, tape, staples, paper clips, etc.
from films before scanning. These may cause serious film
feeding problems and may damage the digitizer's internal
mechanisms. Failure to remove extraneous items from
films will void your warranty.
To prepare films for digitizing:
Stack each group of films according to the sequence specified by the CAD system
manufacturer. A typical sequence is shown below.
Place a separator in each group (not provided by VIDAR). Use separators approved
by the CAD system manufacturer.
VIDAR Advantage Series User's Guide
30cm long x 24cm wide film stack
or
10" long x 12" wide film stack
17