Thumbwheel
5. Tighten the knurled thumbscrew to secure the position.
Complete the Procedure
Run the Benchmark Test diagnostic method (#30) (see
the new syringe works properly.
5.6
Replacing the Air Push Syringe
Parts Needed
•
Air push syringe, 10 mL (P/N 070296)
Center the Syringe Plunger
Follow these steps to center the syringe plunger and prepare the syringe for
removal.
1. Select the Adjust Air Syringe diagnostic method:
a. Press
b. Press
c. Press
Doc. 065330-05 03/2020
Turn thumbwheel one
more partial turn to the
right by the amount shown.
Figure 5-4. Thumbwheel Rotation
to display the
LOAD
LOADING METHOD
to increment the displayed method to 27.
LOAD
once to select method 27.
CONT
Top View
Section
5.17) to verify that
screen.
5 • Service
91