To connect the USB cable:
1. Locate the USB cable (P/N 960777) provided in the Ship Kit
(standard Dionex AutoTrace 280, P/N 071383; Dionex AutoTrace
280 PFAS, P/N 22136-62005).
The flat, rectangular end of the cable is the "A" connector. The
smaller, hexagonal end of the cable is the "B" connector.
2. Plug the "A" connector of the USB cable into a USB 2.0 port on the
PC on which AutoTrace software is installed.
3. Plug the "B" connector of the USB cable into the USB receptacle on
the right side of the Dionex AutoTrace 280 (see
B.3.9

Connecting the Power Cord

Connect a modular power cord (IEC 320 C13) from the main power
receptacle on the Dionex AutoTrace 280 right-side panel to a grounded,
single-phase power source of 100 to 240 Vac, 47/63 Hz.
SHOCK HAZARD—To avoid electrical shock, a grounded receptacle
must be used. Do not operate the Dionex AutoTrace 280 or connect it
to AC power mains without an earthed ground connection.
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the Dionex AutoTrace 280 and is
easily accessible.
Operation at AC input levels outside of the specified operating voltage
range may damage the Dionex AutoTrace 280.
DANGER D'ÉLECTROCUTION—Pour éviter toute électrocution, il faut
utiliser une prise de courant avec prise de terre. Ne l'utilisez pas et ne
le branchez pas au secteur C.A. sans utiliser de branchement mis à la
terre.
Doc. 065330-05 03/2020
NOTE If you connect the USB cable to a USB 3.0 port,
the system will not operate correctly.
B • Installation
Figure
2-5).
143