Making The Required Changes - Thermo Scientific Dionex AutoTrace 280 Operator's Manual

Spe instrument
Hide thumbs Also See for Dionex AutoTrace 280:
Table of Contents

Advertisement

Dionex AutoTrace 280 Operator's Manual
3.7.2

Making the Required Changes

1. Load the method into the Set Up Methods window (see
2. On the toolbar, click Edit.
3. If you want to save the edited method to a new method:
a. Enter a new method name in the Name box.
b. Click the down arrow next to Save, and select Save as new
c. Click Edit to return to editing mode.
4. To edit the settings of an existing step:
a. In the method step table in the lower-right section of the window,
b. Change the settings as required.
5. To delete a step, double-click its entry in the method step table.
6. To add a step:
a. In the method step table, click the step above where you want to
b. In the command list in the left pane of the window, double-click
7. To edit the Dionex AutoTrace 280 operating parameters that will be
in effect when the method is run, click Params on the toolbar. For
details, refer to the "Set Up Method Operating Parameters" topic in
the AutoTrace software Help.
8. To undo changes and restore the previous entries, click Undo on the
toolbar.
9. When changes are complete, click Save on the toolbar.
58
Method.
select the step to be edited. The current settings for the step are
displayed above the table.
add the new step.
the command. The command is added to the table below the
selected step. For details, refer to the "Method Command
Descriptions" topic in the AutoTrace software Help.
NOTE A method can have a maximum of 45 steps.
Section
3.4).
Doc. 065330-05 03/2020