Running Your Method; After Running The Method; Routine Maintenance; Daily Maintenance - Thermo Scientific Dionex AutoTrace 280 Operator's Manual

Spe instrument
Hide thumbs Also See for Dionex AutoTrace 280:
Table of Contents

Advertisement

Dionex AutoTrace 280 Operator's Manual
Load the sample rack.
3.9.2

Running Your Method

Load the method.
Intervene as required.
Press
3.9.3

After Running the Method

Remove the used cartridges or disks.
Prepare to run the next batch of samples by cleaning the sample lines
(see
If done for the day, clean up by emptying the waste containers and
flushing the solvent lines (see
Section

3.10 Routine Maintenance

This section describes routine maintenance procedures that users can perform. All
other Dionex AutoTrace 280 maintenance procedures must be performed by
Thermo Fisher Scientific personnel.
In order to test, adjust, or replace components not covered in this section, run the
Dionex AutoTrace 280 diagnostic methods. For details, see

3.10.1 Daily Maintenance

Check the waste containers and empty as needed.
Flush the solvent lines (see
Flush the sample lines (see
Acetone is harmful to the sample pumps and should never be used to
clean the sample path or rinse the sample container.
Do not run pumps dry. Always leave deionized water in the sample
lines when the system is not in use.
64
to silence the end-of-run alarm.
CONT
Section
3.7.6).
3.8.1).
Section
3.7.7) and sample lines (see
Section
Section
3.7.7).
Section
3.7.8).
4.2.
Doc. 065330-05 03/2020