4.2

Dionex AutoTrace 280 Diagnostic Methods

The Dionex AutoTrace 280 includes several diagnostic methods.
describes the diagnostic methods and explains when to run them.
To run a diagnostic method:
1. Press
LOAD
2. Press
LOAD
example, the number of the Prime Solvents diagnostic method is 29.
3. Press
CONT
4. Press
CONT
Initialization
Syringes
Valves
70
to display the
LOADING METHOD
to increment the displayed method to the required number; for
once to select the method number currently displayed.
to run the selected method.
When you want to...
Initialize system
components (send
syringes to the home
position, etc.)
Diagnose the reason
the system failed to
initialize properly
Check the integrity
of the liquid
handling syringe
drive mechanism
Calibrate the syringe
home position
Check the integrity
of the air push
syringe drive
mechanism
Calibrate the syringe
home position
Confirm the
positioning integrity
of the 12-port valve
Table 4-1. Dionex AutoTrace 280 Diagnostic Methods
Table 4-1
screen.
Run this diagnostic method...
#25: Initialization
#26: Adjust Liq Syringe
#27: Adjust Air Syringe
#28: 12-Port Valve Test
NOTE: Do not run the 12-Port
Valve Test unless all solvent lines
have been primed.
Doc. 065330-05 03/2020