5.10 Calibrating a Sample Pump

Do not calibrate a sample pump unless the volume removed from the sample
container does not match the sample volume requested in the Load Sample step
of your method.
Parts and Tools Needed
Allen wrench, 3/32-in
Sample container with approximately 120 mL of water/pump being calibrated
Balance
1.6-mm (1/16-in) ID tubing (P/N 070358) provided in the Ship Kit (standard
Dionex AutoTrace 280, P/N 071383; Dionex AutoTrace 280 PFAS,
P/N 22136-62005)
Calibrate the Pump
Perform the following procedure on each sample pump that requires calibration.
1. Run a single-step Clean Sample Path method (see
deionized water.
2. Select the Calibrate Pump diagnostic method:
a. Press
b. Press
c. Press
3. Press
CONT
4. Follow the on-screen instructions as they appear.
5. As instructed, weigh the collection vessel before and after pumping.
Determine the percentage difference between the expected volume (100 mL)
and the delivered volume.
For example, if the method pumps 103 mL instead of 100 mL, the volume is
off by 3%.
6. Locate the pump's calibration screw in the hole on the right side of the pump
(see
Figure
Doc. 065330-05 03/2020
to display the
LOAD
LOADING METHOD
LOAD
to increment the displayed method to 31.
once to select method 31.
CONT
to load the method.
5-12).
5 • Service
Section
3.2.2), using
screen.
103