5.5
Replacing the Liquid Handling Syringe
If the liquid handling syringe leaks, replace the syringe.
Parts Needed
•
Liquid handling syringe, 10 mL (P/N 070579)
Purge the Solvent Lines
Follow these steps to purge the liquid handling syringe and its tubing.
NOTE If the syringe is not operable, skip this section and go
1. Remove the solvent lines from their reservoirs.
2. Select the Prime Solvents diagnostic method:
a. Press
b. Press
c. Press
3. Press
CONT
Center the Syringe
Follow these steps to center the syringe plunger and prepare the syringe for
removal.
NOTE If the syringe is not operable, skip this section and go
1. Select the Adjust Liq Syringe diagnostic method:
a. Press
b. Press
c. Press
2. Press
CONT
Doc. 065330-05 03/2020
directly to
"Remove the Defective Syringe" on page
to display the
LOAD
LOADING METHOD
to increment the displayed method to 29.
LOAD
once to select method 29.
CONT
to run the method.
directly to
"Remove the Defective Syringe" on page
to display the
LOAD
LOADING METHOD
LOAD
to increment the displayed method to 26.
once to select method 26.
CONT
to run the method.
5 • Service
88.
screen.
88.
screen.
87